 Network Analyzers using TRVNA software

SCPI Command

SENSe:CORRection:IMPedance[:INPut][:MAGNitude] <numeric>

SENSe:CORRection:IMPedance[:INPut][:MAGNitude]?

Description

Sets or reads out the system impedance Z0 of all Analyzer ports.

command/query

Target

Analyzer

Parameter

<numeric> the Z0 value from 0.001 to 1000.

Unit

Ω (Ohm)

Out of Range

Sets the value of the limit, which is closer to the specified value.

Query Response

<numeric>

Preset Value

50 Ω

Equivalent Softkeys

Calibration > System Z0


Equivalent COM Command

SCPI.SENSe(Ch).CORRection.IMPedance.INPut.MAGNitude

Syntax

Value = app.SCPI.SENSe.CORRection.IMPedance.INPut.MAGNitude

app.SCPI.SENSe.CORRection.IMPedance.INPut.MAGNitude = 50

Type

Double (read/write)


 

Back to SENSe

Back to Command Finder

Back to Command Finder (From User Interface)

 

Rev.:  24.1