SCPI Command
SENSe<Ch>:CORRection:COLLect:ECAL:CHECK:EXECute
Executes the confidence check of the calibration coefficients of the specified channel using the AutoCal module. The command sets the AutoCal Module to the special internal state, reads the S-parameters of this state from the AutoCal Module and sets memory traces so that they can be compared with actual measured data. Comparison is carried out visually by the user. |
---|
no query |
AutoCal module
(<Ch> is arbitrary number 1 to 9)
Calibration > AutoCal > Confidence Check
Equivalent COM Command
SCPI.SENSe(Ch).CORRection.COLLect.ECAL.CHECK.Execute
app.SCPI.SENSe(Ch).CORRection.COLLect.ECAL.CHECK.Execute
Method
Back to SENSe
Back to Command Finder