Please enable JavaScript to view this site.

 Network Analyzers using TRVNA software

SCPI Command

SENSe:CORRection:COLLect:CKIT:ORDer:THRU? <numeric>

Description

Reads out the number of the THRU standard in the calibration kit used to measure between the port 1 and port 2.

query only

Target

Selected calibration kit

Parameter

<numeric>

The number of the calibration standard

Out of Range

If the specified standard number is greater than the number of standards in the kit, an error occurs. If the specified standard number is not the THRU standard number, an error occurs.

Query Response

<numeric>

Equivalent Softkeys

None


Equivalent COM Command

None


 

Back to SENSe

Back to Command Finder

Back to Command Finder (From User Interface)

 

Rev.:  24.1