Calibration and Calibration Kits > Automatic Calibration Module > Automatic Calibration Procedure

One/Two-Port Calibration Procedure

Scroll Contact Us Home Prev Top Next More

For a one/two-port calibration, connect any ports of the ACM to the ports to be calibrated on the Analyzer.

AutoCal ModuleFull 1-port Calibration

To perform full one-port calibration, click the Full 1-Port Calibration softkey.

Wait until calibration is complete.

During the measurement standards, pop-up windows will appear in the channel window. They will have a Calibration label and will indicate the progress of the measurement of the standards.

The instrument status bar will indicate Calibration in progress... when the automatic calibration is in progress.

 

One path 2-port Calibration

To perform one path two-port calibration, click the One Path 2-Port Calibration softkey.

Wait until calibration is complete.

During the measurement standards, pop-up windows will appear in the channel window. They will have a Calibration label and will indicate the progress of the measurement of the standards.

The instrument status bar will indicate Calibration in progress... when the automatic calibration is in progress.

 

 

 

 

 