Calibrations of a four-port Analyzer using a four-port ACM are as simple and automated as possible. For a three-port calibration, connect any three ports of the ACM to the ports to be calibrated on the Analyzer.
For a four-port calibration, all ports of the ACM are connected to the ports of the Analyzer.
To perform full three-port calibration, use the following softkeys: Calibration > AutoCal > 3-Port AutoCal Select the port pair to be calibrated using Select Ports softkey. Click 4-Port Auto Cal Module softkey to start the calibration. Wait until calibration is complete. The instrument status bar will indicate Calibration in progress... when the measurement is in progress. |
To perform full four-port calibration, use the following softkeys: Calibration > AutoCal > 4-Port AutoCal Click 4-Port Auto Cal Module softkey to start the calibration. Wait until calibration is complete. The instrument status bar will indicate Calibration in progress... when the measurement is in progress. |
|
|