Please enable JavaScript to view this site.

 Network Analyzers using S4VNA software

Cable correction function allows to consider the influence of cable characteristics during transform in the time domain. The function contains the cable velocity factor and the cable loss in dB/m. The cable loss value is indicated at the specified frequency. All values can be set manually or selected from the table of predefined cables. The velocity factor is used to convert the time units to the distance units. The cable loss value, together with the frequency, are used to compensate for the attenuation in the cable, so that, for example, the response to an open circuit is unity. The cable correction function is disabled by default.

Cable Correction Activation

Time Domain softkey

Cable Correction

To enable/disable cable correction function of the time domain transformation function, use the following softkeys:

Analysis > Time Domain > Cable Correction > Cable Correction

 

 

 

Cable Table

The software contains the predefined table of cables (See figure below). Each row of the table contains the cable name and the following parameters: velocity factor, cable loss and frequency.

All table fields can be edited. Changes are saved automatically.

If there is no cable description in the table, it is possible to add it. To do this, create a new row in the table using the Add New Cable button and enter its name and parameters.

Cable Correction Table

Cable Table

Select Cable

To open the cable table, use the following softkeys:

Analysis > Time Domain > Cable Correction > Select Cable

Select check softkey

To select the cable in table, use the Select softkey.

Note: Make sure that the selected cable is check marked.

Add new cable

Delete cable

To add the new cable in the table, use the Add New Cable softkey:

Note: A new cable can be added in the table by specifying its name and parameters in the empty field at the end of the table.

To delete cable table, use the Delete Cable softkey.

Restore Table

To restore cable table, use the following softkeys:

Restore Cable > ОК

Save Load cable table 

To save cable table in file, use the Save Cable Table to File... softkey.

To load cable table in file, use the Load Cable Table from File... softkey.

Velocity Factor

Velocity factor is used to calculate distance along a cable from the cable delay value. If the cable correction function is disabled, the software assumes it to be equal to 1. To obtain the accurate mismatch location in a cable, it is important to set the right velocity factor of the cable.

Velocity factor

To set the velocity cable, use the following softkeys:

Analysis > Time Domain > Cable Correction > Velocity Factor

 

 

 


note

The velocity factor value can also be set by the selecting the cable in the cable table.


Cable Loss

The cable loss value is used to compensate the signal attenuation in a cable. The cable loss value is set in dB per meter.

Cable Loss

To set the cable loss, use the following softkeys:

Analysis > Time Domain > Cable Correction > Cable Loss

 

 

 


note

The cable loss value can also be set  by the selecting the cable in the cable table.


Frequency

Fequency

To set the frequency, at which the cable loss is specified, use the following softkeys:

Analysis > Time Domain > Cable Correction > Frequency

 

 

 

 

Rev.:  22.4