Please enable JavaScript to view this site.

 Network Analyzers using S2VNA software

The function sets the trigger source to start measuring the calibration standards. If an Internal source is selected, the calibration starts immediately. If the source is System, the system trigger is used to start the calibration. The source of the system trigger is set by the softkey:

Stimulus > Trigger > Trigger Source > [Internal | External | Manual | Bus]

If a system trigger is used, the averaging trigger function and the external trigger event function set On Point affect the start of the calibration in the same way as during standard measurements. When using a system trigger, the trigger source Bus should not be used, otherwise the software may be blocked.


note

This function does not apply to calibration using the ACM, power calibration, receiver calibration. In these calibrations, the internal trigger is always used.


Calibration

Cal Trig Source

To set the calibration trigger source, use the following softkeys:

Calibration > Cal Trig Source [Internal | System]

 

 

 

 

Rev.:  24.2