This setting allows to select the external trigger event.
Trigger event |
Function |
---|---|
On sweep [default] |
One trigger signal starts a full measurement cycle, that is, the measurement of all frequency points of all channels included in the measurement cycle. |
One trigger signal starts the measurement of one frequency point of a channel. The next trigger signal starts the measurement of the next frequency point of the channel, and so on. |
note |
If the Averaging Trigger function and the On point trigger function are enabled at the same time, the On point trigger function has priority. In this case, N * P trigger signals are required to complete the averaging, where N is the averaging factor, and P is the number of points. |
To select an external trigger event, use the following softkeys: Stimulus > Trigger > Ext Trigger > Event Then select the required external trigger event: •On Sweep •On Point |
|
|