SCPI Command
CALCulate<Ch>[:SELected]:LIMit:DATA <numeric list>
CALCulate<Ch>[:SELected]:LIMit:DATA?
Description
Sets the data array, which is the limit line in the limit test function. The array size is 1 + 5N, where N is the number of limit line segments. For the n–th point, where n from 1 to N: <numeric 1> the number of limit line segments N is from 0 to 100. Setting 0 clears the limit line. <numeric 5n–3> type of the n–th limit line segment: 0: Off. 1: Upper limit 2: Lower limit 3: Single Point limit <numeric 5n–2> the stimulus value in the start point of the n-th segment <numeric 5n–1> the stimulus value in the end point of the n-th segment <numeric 5n–0> the response value in the start point of the n-th segment <numeric 5n+1> the response value in the end point of the n-th segment Note: If the array size is not 1 + 5N, where N is <numeric 1>, an error occurs. If <numeric 5n–3> is less than 0 or more than 2, an error occurs. When <numeric 5n–2>, <numeric 5n–1>, <numeric 5n–0>, and <numeric 5n+1> elements are out of allowable range, the value is set to the limit, which is closer to the specified value. |
---|
command/query |
The active trace of channel <Ch>,
<Ch>={[1]|2|3|4}
<Ch>={[1] … |16} (in N-port mode only)
<numeric 1>, <numeric 2>, …<numeric 5N+1>
Equivalent Softkeys
Analysis > Limit Test > Edit Limit Line
Equivalent COM Command
SCPI.CALCulate(Ch).SELected.LIMit.DATA
Data = app.SCPI.CALCulate(Ch).SELected.LIMit.DATA
app.SCPI.CALCulate(Ch).SELected.LIMit.DATA = Array(1,2,800,900,–10,–10)
Variant (array of Double) (read/write)
Back to CALCulate
Back to Command Finder
Back to Command Finder (From User Interface)