Please enable JavaScript to view this site.

 Network Analyzers using PXI software

SCPI Command

SENSe<Ch>:OFFSet:PORT<Pt>[:FREQuency]:OFFSet <frequency>

SENSe<Ch>:OFFSet:PORT<Pt>[:FREQuency]:OFFSet?

Description

Sets or reads out the basic frequency range offset of port <Pt> when the frequency offset feature is ON and offset type is "PORT".

command/query

Target

Port <Pt> of channel <Ch>,

<Ch>={[1]|2|...16}

<Pt>={[1]|2}

Parameter

<frequency> offset from –1e12 to 1e12

Unit

Hz (Hertz)

Out of Range

Sets the value of the limit, which is closer to the specified value.

Query Response

<numeric>

Preset Value

0

Related Commands

SENS:OFFS

SENS:OFFS:TYPE

Equivalent Softkeys

Stimulus > Frequency Offset > Port n > Offset

Back to SENSe

Back to Command Finder

Back to Command Finder (From User Interface)

 

Rev.:  22.3