SCPI Command
SENSe<Ch>:CORRection:OFFSet:COLLect:SAVE
Calculates the calibration coefficient for the selected calibration type (scalar mixer calibration only) from the calibration data measured with the frequency offset feature is ON. On completion of the command, all the calibration standards measurements are cleared, and the error correction automatically turns ON. If this command is executed before all necessary calibration data for calculating the calibration coefficient is measured, an error occurs when executed. |
---|
no query |
Channel <Ch>,
<Ch>={[1]|2|...16}
SENS:CORR:OFFS:COLL:METH:SMIX2
Calibration > Mixer/Converter Calibration > Scalar Mixer Calibration > Apply
Back to SENSe
Back to Command Finder
Back to Command Finder (From User Interface)