SCPI Command
SENSe<Ch>:CORRection:COLLect:THRU:ADDition:PERMittivity <numeric>
SENSe<Ch>:CORRection:COLLect:THRU:ADDition:PERMittivity?
Sets or reads out the value of the permittivity of the thru media in the thru addition function. This parameter is used to calculate the adapter delay when the thru length is setting; therefore, this parameter must be set before setting of the thru length. |
---|
command/query |
Channel <Ch>,
<Ch>={[1]|2|...16}
<numeric> the value of the permittivity of the thru
<numeric>
1.000649 (air)
Calibration > Calibrate > Thru Addition > Permittivity
Back to SENSe
Back to Command Finder
Back to Command Finder (From User Interface)