Please enable JavaScript to view this site.

 Network Analyzers using PXI software

SCPI Command

SENSe<Ch>:CORRection:COLLect:DATA:OPEN <port>,<numeric list>

SENSe<Ch>:CORRection:COLLect:DATA:OPEN? <port>

Description

Writes or reads out the array of the open calibration standard measurement for the port <port>.

The array size is 2N, where N is the number of measurement points.

For the n–th point, where n from 1 to N:

<numeric 2n–1> real part of the measurement

<numeric 2n> imaginary part of the measurement

command/query

Target

Channel <Ch>,

<Ch>={[1]|2|...16}

Parameter

<port>

The number of the port from 1 to 4

<numeric list>

The data array of the open standard measurement

Query Response

<numeric 1>, <numeric 2>, …<numeric 2N>

Related Commands

SENS:CORR:COLL:OPEN

Equivalent Softkeys

None

Back to SENSe

Back to Command Finder

Back to Command Finder (From User Interface)

 

Rev.:  22.3