SCPI Command
SENSe<Ch>:CORRection:COLLect:DATA:OPEN <port>,<numeric list>
SENSe<Ch>:CORRection:COLLect:DATA:OPEN? <port>
| Writes or reads out the array of the open calibration standard measurement for the port <port>. The array size is 2N, where N is the number of measurement points. For the n–th point, where n from 1 to N: <numeric 2n–1> real part of the measurement <numeric 2n> imaginary part of the measurement | 
|---|
| command/query | 
Channel <Ch>,
<Ch>={[1]|2|...16}
| <port> | The number of the port from 1 to 4 | 
| <numeric list> | The data array of the open standard measurement | 
<numeric 1>, <numeric 2>, …<numeric 2N>
None
Back to SENSe
Back to Command Finder