SCPI Command
SENSe<Ch>:CORRection:COLLect:ADAPter:LENGth <numeric>
SENSe<Ch>:CORRection:COLLect:ADAPter:LENGth?
Sets or reads out the approximate value of the mechanical length of the adapter in the adapter removal/insertion function. This value is used to eliminate the uncertainty of ± 180 ° when calculating the phase response of the adapter. The sign of the value depends on the type of the removal / insertion function. The value must be negative for the adapter removal function and must be positive for the adapter insertion function. If this value is set to zero, the analyzer uses an algorithm to automatically determine the delay of the adapter. In most cases setting this value to zero is enough. Setting this value to non-zero is required when:
Note: The delay and the length of the adapter can be set mutually |
---|
command/query |
Channel <Ch>,
<Ch>={[1]|2|...16}
<numeric> the approximate delay length of the adapter (with minus sign when adapter is removed).
m (Meters)
<numeric>
0
Calibration > Calibrate > Adapter Removal / Insertion > Adapter Delay
Back to SENSe
Back to Command Finder
Back to Command Finder (From User Interface)