Please enable JavaScript to view this site.

 CMT VNA Programming Manual

SCPI Command

SENSe<Ch>:CORRection:COLLect:METHod[:RESPonse]:OPEN <port>

Description

Selects the port and sets the response calibration (Open) type for the calculation of the calibration coefficients on completion of the calibration executed by the SENS:CORR:COLL:SAVE command.

no query

Target

Channel <Ch>,

<Ch>={[1]|2|...16}

Parameter

<port>

The number of the port from 1 to 2

Related Commands

SENS:CORR:COLL:SAVE

Back to SENSe

Back to Command Finder

 

Rev.:  26.00