 Network Analyzers using TRVNA software

The S-parameter conversion function allows the conversion of measurement results (S11 or S21) to the following parameters:

Parameter

Equation

Impedance in reflection measurement ()

Admittance in reflection measurement ()

Impedance in transmission measurement ()

Admittance in transmission measurement ()

Inverse S-parameter for reflection and transmission measurements respectively

 and

S-parameter complex conjugate

 and

is characteristic impedance of port 1.

is characteristic impedance of port 2.

and are measured S-parameters (1 and 2 are the port identifiers).


note

Equations for , , , are approximate. The reason for using the approximate method is the measurement speed, as only one S-parameter is used in the calculations.


The S-parameter conversion function can be applied to an individual trace of a channel. The trace to which the function is applied must be preselected as active (See Selection of Active Trace/Channel).

Conversion softkey

Conversion OFF

To enable/disable the conversion, use the following softkeys:

Analysis > Conversion > Conversion

 

Conversion function

To select the conversion type, use the following softkeys:

Analysis > Conversion > Function

Then select the required function:

Impedance Z

Admittance Y

Inverse 1/S

Conjugation

 

 

 


note

All conversion types are indicated in the trace status field, when enabled.


 