note |
This section is available for TR1300/1 only. The TR1300/1 analyzer operates with an internal reference frequency (10 MHz) oscillator only. |
The frequency reference offset function can be used if a more accurate reference frequency is required. Connect an external frequency counter to the 10MHz Reference Output connector on the Analyzer's rear panel. Measure the actual value of the reference frequency. Enable the reference offset function and enter the measured reference frequency value. The deviation of the reference frequency will be compensated.
To enable the reference offset function, use the following softkeys: System > Misc Setup > Reference Offset [ON | OFF] |
To enter the value measured by the frequency counter, use the following softkeys: System > Misc Setup > Measured Ref. |
|
|