The bandwidth search function allows to determine and view the following parameters of a passband or a stopband: bandwidth, center frequency, lower frequency, higher frequency, Q value, and insertion loss (See figure below).
The bandwidth search is executed from the reference point. The active marker or the maximum trace value can be selected as the reference. The bandwidth search function detects lower and higher cutoff frequencies that differ from the reference point response by a user-specified bandwidth value (usually – 3 dB).
Parameter Description |
Symbol |
Definition |
Formula |
---|---|---|---|
Bandwidth |
BW |
The difference between the higher and lower cutoff frequencies. |
F2 – F1 |
Center Frequency |
cent |
The midpoint between the higher and lower cutoff frequencies. |
(F1+F2)/2 |
Lower Cutoff Frequency |
low |
The lower frequency point of the intersection of the bandwidth cutoff level and the trace. |
F1 |
Higher Cutoff Frequency |
high |
The higher frequency point of the intersection of the bandwidth cutoff level and the trace. |
F2 |
Quality Factor |
Q |
The ratio of the center frequency to the bandwidth. |
cent/BW |
Loss |
loss |
The trace measured value in the reference point of the bandwidth search. |
— |
To enable/disable bandwidth search function, use the following softkeys: Markers > Marker Math > Bandwidth Search > Bandwidth Search |
Set the bandwidth search type by softkeys: Markers > Marker Math > Bandwidth Search > Type The type and the softkey label toggle between Bandpass and Notch settings. |
To set the search reference point, use the following softkeys: Markers > Marker Math > Bandwidth Search > Search Ref To > [Marker | Maximum | Minimum] |
To enter the bandwidth value, use the following softkeys: Markers > Marker Math > Bandwidth Search > Bandwidth Value |
|
|