SCPI Command

CALCulate<Ch>[:SELected]:MARKer<Mk>:BWIDth:DATA?

Description

Reads out the bandwidth search result.

The bandwidth search can be performed relatively to the marker <Mk>, or relatively to the absolute maximum value of the trace (in this case the number of the marker is ignored), what is set by the CALC:MARK:BWID:REF command.

The data include 4 elements:

<numeric 1>      Bandwidth

<numeric 2>      Center frequency

<numeric 3>      Q value

<numeric 4>      Loss

Note: If the bandwidth search is impossible, all the read-out values are 0. If the search is performed relatively to a maker, which is OFF, an error occurs.

query only

Target

Marker <Mk> of the active trace of channel <Ch>,

<Ch>={[1]|2|...9}

<Mk>={[1] … |16}

Query Response

<numeric 1>, <numeric 2>, …<numeric 4>

Related Commands

CALC:MARK:BWID:REF

Equivalent Softkeys

None


Equivalent COM Command

SCPI.CALCulate(Ch).SELected.MARKer(Mk).BWIDth.DATA

Syntax

Data = app.SCPI.CALCulate(Ch).SELected.MARKer(Mk).BWIDth.DATA

Type

Variant (array of Double) (read only)


 