Please enable JavaScript to view this site.

 Network Analyzers using S4VNA software

 


note

The availability of the TRL calibration depends on the Analyzer model (See corresponding datasheet).


Here is an example of calibration using the calibration kit for TRL calibration, in which the "TRL LINE/MATCH" class contains 3 subclasses: load (Lowband), line 2 (TRL Line2), and line 3 (TRL Line3).

1-2 Line_Match 05CK120-150

Subclass 1

Subclass 2 Line2

Subclass 3 Line3

In the main menu of TRL calibration the 1-2 Line/Match softkey will open the subclass menu (if the above mentioned condition is met).

Connect the Lowband, Line2 and Line3 to the test ports in any consequence and perform measurements clicking the softkey corresponding to the connected standard.

Info softkey

If two standards have an overlapping frequency range, the last measured standard will be used in the overlapping region.

To view additional information about each standard frequency range, in which its measurements are applied (See figure below), press the Info softkey.

Information on calibration standard measurements

Information on calibration standard measurements

 

Rev.:  22.4