 Network Analyzers using S4VNA software

Before calibrating the Analyzer with the ACM, perform some settings, i.e. activate a channel and set channel parameters (frequency range, IF bandwidth, etc.).

Connect the ACM to the Analyzer test ports and connect the USB port of the ACM to the USB port of the PC.

AutoCal Module

To open automatic calibration submenu, use the following softkeys:

Calibration > AutoCal

Select Characterization

To select characterization, use the following softkeys:

Characterization > [Factory | User1 | User2 | User3]

 

Select Orientation 4-port

When selecting manual or automatic orientation for ACM, it is recommended to select the automatic orientation.

To enable auto orientation for ACM before performing each automatic calibration, use the following softkeys:

Orientation > Auto-Orientation [ON | OFF]

To manually select the orientation, turn OFF Auto-Orientation. Use the Port n softkeys to relate the analyzer test ports with the ACM ports.

Regardless of the other settings, automatic orientation is performed immediately by clicking the Execute Auto-Orientation softkey.

 

Unkn Thru OFF

Enable or disable Unknown Thru algorithm using the Unkn Thru [ON | OFF] softkey.

 

Thermo compensation ON

Enable or disable the thermal compensation using the Thermo compensation [ON | OFF] softkey.

Characterization Info...

To display detailed information on characterization, use the Characterization Info softkey.

 

 

 

 

 

Rev.:  22.4