SCPI Command
SENSe<Ch>:SEGMent:DATA <numeric list>
SENSe<Ch>:SEGMent:DATA?
Sets or reads out the array of the segment sweep table. The array has the following format: {<Cnt>, <Flag1>, <Flag2>, <Flag3>, <Flag4>, <Flag5>[,<Flag6], <N>, <Start 1>, <Stop 1>, <NOP 1> [,<IFBW 1>] [,<Pwr 1>] [,<Del 1>] [,<Time 1>][,<AuxPwr 1>], <Start 2>, <Stop 2>, <NOP 2> [,<IFBW 2>] [,<Pwr 2>] [,<Del 2>] [,<Time 2>][,<AuxPwr 2>], … <StartN>, <StopN>, <NOP N> [,<IFBW N>] [,<Pow N>] [,<Del N>] [,<TimeN>][,<AuxPwr N>]}
<Cnt> : 5 - specifies <Flag1> to <Flag5> 6 - specifies <Flag1> to <Flag6> <Flag1> : Stimulus mode setting (0 — start/stop, 1 — center/span), <Flag2> : Setting of the <IFBW> field (0 — disabled, 1 — enabled), <Flag3> : Setting of the <Pwr> field (0 — disabled, 1 — enabled), <Flag4> : Setting of the <Del> field (0 — disabled, 1 — enabled), <Flag5> : Setting of the <Time> field (0 — disabled, 1 — enabled), <Flag6> : Setting of the <AuxPwr> field (0 — disabled, 1 — enabled), <N> : Number of segments, <Start n> : Start value of the n-th segment, <Stop n> : Stop value of the n-th segment, <NOP n> : Number of points of the n-th segment, <IFBW n> : IF bandwidth of the n-th segment (if enabled), <Pwr n> : Power of the n-th segment (if enabled), <Del n> : Measurement delay of the n-th segment (if enabled), <Time n> : Reserved for future use (if enabled), <AuxPwr n> : Auxiliary Source Power of the n-th segment (if enabled) |
---|
command/query |
Channel <Ch>,
<Ch>={[1]|2|...16}
<numeric 1>,<numeric 2>,…<numeric (Cnt+2)+M×N>
Where:
N – the number of the segments,
M – depends on the values of the flags:
M = 3 + <Flag2> + <Flag3> + <Flag4> + <Flag5>+<Flag6>
Stimulus > Segment Table
Equivalent COM Command
SCPI.SENSe(Ch).SEGMent.DATA
Data = app.SCPI.SENSe(Ch).SEGMent.DATA
app.SCPI.SENSe(Ch).SEGMent.DATA = Data
Variant (array of Double) (read/write)
Back to SENSe
Back to Command Finder
Back to Command Finder (From User Interface)