Please enable JavaScript to view this site.

 Network Analyzers using S4VNA software

SCPI Command

SENSe<Ch>:CORRection:RECeiver<Pt>:COLLect:RCHannel:ACQuire <srcport>

Description

Executes receiver calibration of the reference receiver of the specified port <Pt>. The test receiver calibration uses port number <srcport> as the source port.

Note: The source port number must be the same as the receiver port number, except for VNAs with direct access to the receiver.

Note: The command starts the measurement of the calibration data immediately regardless the trigger settings. The command blocks the execution of the subsequent commands until the completion of the measurement.

no query

Target

Port <Pt> of channel <Ch>,

<Ch>={[1]|2|...16}

<Pt>={[1]|2|3|4}

<Pt>={[1]|2|3|4}

Parameter

<srcport>

The number of the source port from 1 to 4

Equivalent Softkeys

Calibration > Receiver Calibration > Calibrate Reference Receiver


Equivalent COM Command

SCPI.SENSe(Ch).CORRection.RECeiver(Pt).COLLect.RCHannel.ACQuire

Syntax

app.SCPI.SENSe(Ch).CORRection.RECeiver(Pt).COLLect.RCHannel.ACQuire

Type

Method


 

Back to SENSe

Back to Command Finder

Back to Command Finder (From User Interface)

 

Rev.:  22.4