Please enable JavaScript to view this site.

 Network Analyzers using S4VNA software

SCPI Command

SENSe<Ch>:CORRection:OFFSet:COLLect:SAVE

Description

Calculates the calibration coefficient for the selected calibration type (scalar mixer calibration only) from the calibration data measured with the frequency offset feature is ON.

On completion of the command, all the calibration standards measurements are cleared, and the error correction automatically turns ON.

If this command is executed before all necessary calibration data for calculating the calibration coefficient is measured, an error occurs when executed.

no query

Target

Channel <Ch>,

<Ch>={[1]|2|...16}

Related Commands

SENS:CORR:OFFS:COLL:METH:SMIX2

SENS:CORR:OFFS:COLL:LOAD

SENS:CORR:OFFS:COLL:OPEN

SENS:CORR:OFFS:COLL:SHOR

SENS:CORR:OFFS:COLL:THRU

SENS:CORR:OFFS:COLL:PMETer

Equivalent Softkeys

Calibration > Mixer/Converter Calibration > Scalar Mixer Calibration > Apply


Equivalent COM Command

SCPI.SENSe(Ch).CORRection.OFFSet.COLLect.SAVE

Syntax

app.SCPI.SENSe(Ch).CORRection.OFFSet.COLLect.SAVE

Type

Method


 

Back to SENSe

Back to Command Finder

Back to Command Finder (From User Interface)

 

Rev.:  22.4