SCPI Command
SENSe<Ch>:CORRection:OFFSet:COLLect[:ACQuire]:ECAL <numeric1>,<numeric2>
Measures the calibration data of all reflection standards of the ACM on the specified port when the frequency offset feature is on for scalar mixer calibration. Use ACM measurement instead of measuring three mechanical standards: OPEN, SHORT, LOAD. Note: The command starts the measurement immediately independent on the trigger settings. |
---|
no query |
Channel <Ch>,
<Ch>={[1]|2|...16}
<numeric1> |
Measurement port number |
<numeric2> |
Number of the second port of the SMC port pair |
If an incorrect port number is specified, an error occurs.
Calibration > Mixer/Converter Calibration > Scalar Mixer Calibration > Measure Using ACM
Equivalent COM Command
None