SCPI Command
SENSe<Ch>:CORRection:EXTension:AUTO:MEASure {SHORt|OPEN}
Performs measurement of the standard "SHORT" or "OPEN", automatically calculates and sets the parameters of the port extension. The set of ports for which this command is executed is determined by the SENS:CORR:EXT:AUTO:PORT command. When two consecutive measurements of "SHORT" and "OPEN" are performed the results of these measurements are averaged. |
---|
command |
Channel <Ch>,
<Ch>={[1]|2|...16}
SHORt |
Measures "SHORT" standard |
OPEN |
Measures "OPEN" standard |
Calibration > Port Extension > Auto Port Extension > Measure Short | Measure Open
Equivalent COM Command
None
Back to SENSe
Back to Command Finder
Back to Command Finder (From User Interface)