Please enable JavaScript to view this site.

 Network Analyzers using S4VNA software

SCPI Command

SENSe<Ch>:CORRection:COLLect:THRU:ADDition:FULL2:COMPlete <port1>,<port2>

Description

Completes the full two-port calibration between the specified ports provided that each port was calibrated using full one-port calibration:

Measures an unknown thru between the ports.

Calculates the error terms Et and El using the unknown thru algorithm.

Saves the Et and El error terms to the existing calibration getting the full two-port calibration from the two one-port calibrations.

If the full two-port calibration already existed between the specified ports, updates the Et and El error terms.

no query

Target

Channel <Ch>,

<Ch>={[1]|2|...16}

Parameter

<port1>

The first port number from 1 to 4

<port2>

The second port number from 1 to 4

Equivalent Softkeys

Calibration > Calibrate > Thru Addition > Complete 2-Port Calibration


Equivalent COM Command

None


 

Back to SENSe

Back to Command Finder

Back to Command Finder (From User Interface)

 

Rev.:  22.4