Please enable JavaScript to view this site.

 Network Analyzers using S4VNA software

SCPI Command

SENSe<Ch>:CORRection:COLLect:METHod:TYPE?

Description

Reads out the calibration method selected for the calculation of the calibration coefficients on completion of the calibration executed by the SENS:CORR:COLL:SAVE command.

query only

Target

Channel <Ch>,

<Ch>={[1]|2|...16}

Query Response

RESPO

Response (Open)

RESPS

Response (Short)

RESPT

Response (Thru)

SOLT1

Full one-port calibration

SOLT2

Full two-port calibration

1PATH

One path two-port calibration

NONE

Not defined

Equivalent Softkeys

None


Equivalent COM Command

SCPI.SENSe(Ch).CORRection.COLLect.METHod.TYPE

Syntax

Param = app.SCPI.SENSe(Ch).CORRection.COLLect.METHod.TYPE

Type

String (read only)>


 

Back to SENSe

Back to Command Finder

Back to Command Finder (From User Interface)

 