SCPI Command
SENSe<Ch>:CORRection:COLLect:METHod:ERESponse <rcvport>,<srcport>
Selects the ports and sets the one path 2–port calibration type for the calculation of the calibration coefficients on completion of the calibration executed by the SENS:CORR:COLL:SAVE command. |
---|
no query |
Channel <Ch>,
<Ch>={[1]|2|...16}
<rcvport> |
The number of the receiver port from 1 to 4 |
<srcport> |
The number of the source port from 1 to 4 |
If the same port numbers are specified, an error occurs.
Calibration > Calibrate > One Path 2–Port Cal > Select Port
Equivalent COM Command
SCPI.SENSe(Ch).CORRection.COLLect.METHod.ERESponse
app.SCPI.SENSe(Ch).CORRection.COLLect.METHod.ERESponse = Array(2, 1)
Variant (array of Long) (write only)