SCPI Command
SENSe:CORRection:COLLect:CKIT:STAN<Std>:Z0 <numeric>
SENSe:CORRection:COLLect:CKIT:STAN<Std>:Z0?
Sets or reads out the offset Z0 value for the calibration standard. |
---|
command/query |
Standard <Std> of the calibration kit,
<Std>={[1]|2|...N}, where N — the number of the standards in the calibration kit
<numeric> the offset Z0 value from –1E18 to 1E18
Ω (Ohm)
Sets the value of the limit, which is closer to the specified value.
<numeric>
50 or 75 Ω, depending on the selected calibration kit
Calibration > Cal Kit > Define STDs -> {Open | Short | Load | Thru/Delay} -> Offset -> Z0
Equivalent COM Command
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).Z0
Value = app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).Z0
app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).Z0 = 50
Double (read/write)
Back to SENSe
Back to Command Finder