SCPI Command

SENSe:CORRection:COLLect:CKIT:STAN<Std>:DATA <numeric list>

SENSe:CORRection:COLLect:CKIT:STAN<Std>:DATA?

Description

Writes or reads out the data array of the data-based calibration standard. The first element of the array is 1 or 2 and determines the number of ports of the calibration standard. The array format is as follows.

When the first element of the array is 1 :

<1>,<freq1>,<S11.re1>,<S11.im1>,

<freq2>,<S11.re2>,<S11.im2>,

<freqN>,<S11.reN>,<S11.imN>

When the first element of the array is 2:

<2>,<freq1>,<S11.re1>,<S11.im1>,<S21.re1>,<S21.im1>,

<S12.re1>,<S12.im1>,<S22.re1>,<S22.im1>,

<freqN>,<S11.reN>,<S11.imN>,<S21.reN>,<S21.imN>,

<S12.reN>,<S12.imN>,<S22.reN>,<S22.imN>

command/query

Target

Standard <Std> of the selected calibration kit,

<Std>={[1]|2|...N}, where N — the number of the standards in the calibration kit

Query Response

<numeric 1>, <numeric 2>, …<numeric N>

Equivalent Softkeys

Calibration > Cal Kit > Define STDs > Define STD Data


Equivalent COM Command

None


 

Back to SENSe

Back to Command Finder