SCPI Command
SENSe<Ch>:CORRection:COLLect:ADAPter:PERMittivity <numeric>
SENSe<Ch>:CORRection:COLLect:ADAPter:PERMittivity?
Sets or reads out the value of the permittivity of an adapter media in the adapter removal/insertion function. When setting the adapter length, this parameter is used to calculate the adapter delay; therefore, this parameter must be set before setting the adapter length. This parameter is not used when setting the adapter delay. |
---|
command/query |
Channel <Ch>,
<Ch>={[1]|2|...16}
Parameter
<numeric> the value of the permittivity of an adapter
<numeric>
1.000649 (air)
Calibration > Calibrate > Adapter Removal / Insertion > Permittivity
Equivalent COM Command
None
Back to SENSe
Back to Command Finder
Back to Command Finder (From User Interface)