SCPI Command
SENSe<Ch>:CORRection:COEFficient:METHod:SOLT1 <port>
Selects the port and sets the full one–port calibration type when the written calibration coefficients are made effective by the SENS:CORR:COEF:SAVE command. |
---|
no query |
Channel <Ch>,
<Ch>={[1]|2|...16}
<port> |
The number of the port from 1 to 4 |
None
Equivalent COM Command
SCPI.SENSe(Ch).CORRection.COEFficient.METHod.SOLT1
Port = 1
app.SCPI.SENSe(Ch).CORRection.COEFficient.METHod.SOLT1= Port
Type
Long (write only)
Back to SENSe
Back to Command Finder