Programming > Connection Setup

Connecting Multiple Analyzers to Single Computer

Scroll Contact Us Home Prev Top Next More

The section describes in detail how to configure remote control of multiple analyzer programs executed simultaneously on a single PC (provided several USB analyzer hardware units connected to the single PC).

It is recommended to create a separate folder for each Analyzer with the software. This allows to save individual settings for each Analyzer.

It is recommended that each copy of the software be linked to a specific hardware unit by its serial number or model (See at the end of this section).

Assign a unique TCP/IP port number for each copy of the software for the HiSLIP or Socket protocol used. For example, if HiSLIP is used, assign port 4880 to the first analyzer, 4881 to the second, and so on. When assigning a port number, the user must ensure that the port number is not in use by other programs.

Use the Analyzer's address in the user program with the mandatory indication of the TCP/IP port number assigned to the Analyzer, as in the examples given.

Examples of the VISA address for the HiSLIP and Socket protocols with the indication of the TCP/IP port

HiSLIP

TCPIP0::192.168.0.1::hislip0,4880::INSTR

TCPIP0::192.168.0.1::hislip0,4881::INSTR

Socket

TCPIP0::192.168.0.1::5025::SOCKET

TCPIP0::192.168.0.1::5026::SOCKET

Examples of the TCP/IP address of the analyzer's Socket server with the indication of the port

Socket

192.168.0.1:5025

192.168.0.1:5026

Misc SetupAnalyzer ModelAnalyzer Serial N

To link the analyzer program to the analyzer model, press the following softkeys:

System > Misc Setup > Analyzer Model

To link the analyzer program to the analyzer serial number, press the following softkeys:

System > Misc Setup > Analyzer Serial N

 