Please enable JavaScript to view this site.

 Network Analyzers using S4VNA software

Typically, the client uses the VISA library to establish connection to analyzer software. The easiest way to configure the network connection with the Analyzer is using a special utility from VISA package (for example, NI-MAX, Keysight Connection Expert).

Following the manual for the above utilities, add a new network device — specifying the network name or IP address of the Analyzer's PC — and the protocol. Once successfully connected to the Analyzer, the VISA address of the Analyzer will be automatically generated and displayed. Use this VISA address in the client program in order to open the connection.

The format of the VISA address for the HiSLIP and Socket protocols

HiSLIP

TCPIP[board]::host address[::HiSLIP device name[,HiSLIP port]][::INSTR]

Socket

TCPIP[board]::host address::port::SOCKET

Examples of VISA address for HiSLIP and Socket protocols

HiSLIP

TCPIP0::192.168.0.1::hislip0::INSTR

TCPIP0::localhost::hislip0::INSTR

Socket

TCPIP0::192.168.0.1::5025::SOCKET

TCPIP0::localhost::5025::SOCKET

If the client is a user program that does not use the VISA library, then the only available protocol is the TCP/IP Socket protocol. In this case, the user program establishes a connection using the IP address of the Analyzer's Socket server.

The format of the IP address of the analyzer's Socket server

Socket

host address:port

Examples of the IP address of the analyzer's Socket server

Socket

192.168.0.1:5025

localhost:5025

 