Please enable JavaScript to view this site.

 Network Analyzers using S4VNA software

SCPI Command

CALCulate<Ch>[:SELected]:TRANsform:TIME:STEP:RTIMe <time>

CALCulate<Ch>[:SELected]:TRANsform:TIME:STEP:RTIMe?

Or

CALCulate<Ch>:TRACe<Tr>:TRANsform:TIME:STEP:RTIMe <time>

CALCulate<Ch>:TRACe<Tr>:TRANsform:TIME:STEP:RTIMe?

Description

Sets or reads out the rise time of the step signal (time domain transformation resolution), coupled with the Kaiser-Bessel window shape β parameter. The impulse width setting changes the β parameter and setting of β parameter changes the impulse width.

command/query

Target

CALCulate<Ch>[:SELected] — active trace of channel <Ch>,

Or

CALCulate<Ch>:TRACe<Tr> — trace <Tr> of channel <Ch>,

<Ch> = {[1]|2|...16}

<Tr> = {[1]|2|...16}

Parameter

<time> the impulse width, the allowable range varies depending on the specified frequency range

Unit

sec (second)

Out of Range

Sets the value of the limit, which is closer to the specified value.

Query Response

<numeric>

Equivalent Softkeys

Analysis > Time Domain > Window > Impulse Width

(when the transformation type is set to Lowpass Step)


Equivalent COM Command

SCPI.CALCulate(Ch).SELected.TRANsform.TIME.RTIMe


note

This command is similar to CALCulate<Ch>[:SELected]:TRANsform:TIME:STEP:RTIMe only.


Syntax

Value = app.SCPI.CALCulate(Ch).SELected.TRANsform.TIME.RTIMe

app.SCPI.CALCulate(Ch).SELected.TRANsform.TIME.RTIMe = 1e–8

Type

Double (read/write)


 

Back to CALCulate

Back to Command Finder

Back to Command Finder (From User Interface)

 

Rev.:  22.4