SCPI Command
CALCulate<Ch>[:SELected]:TRANsform:TIME:STARt <time>
CALCulate<Ch>[:SELected]:TRANsform:TIME:STARt?
Or
CALCulate<Ch>:TRACe<Tr>:TRANsform:TIME:STARt <time>
CALCulate<Ch>:TRACe<Tr>:TRANsform:TIME:STARt?
Sets or reads out the time domain start value when the time domain transformation function is turned ON. |
---|
command/query |
CALCulate<Ch>[:SELected] — active trace of channel <Ch>,
Or
CALCulate<Ch>:TRACe<Tr> — trace <Tr> of channel <Ch>,
<Ch> = {[1]|2|...16}
<Tr> = {[1]|2|...16}
<time> the time domain start value, the allowable range varies depending on the specified frequency range and the number of points
sec (second)
Sets the value of the limit, which is closer to the specified value.
<numeric>
-1e-8
Analysis > Time Domain > Start
Equivalent COM Command
SCPI.CALCulate(Ch).SELected.TRANsform.TIME.STARt
note |
This command is similar to CALCulate<Ch>[:SELected]:TRANsform:TIME:STARt only. |
Value = app.SCPI.CALCulate(Ch).SELected.TRANsform.TIME.STARt
app.SCPI.CALCulate(Ch).SELected.TRANsform.TIME.STARt = 1e–8
Double (read/write)
Back to CALCulate