SCPI Command
CALCulate<Ch>[:SELected]:TRANsform:TIME:IMPulse:WIDTh <time>
CALCulate<Ch>[:SELected]:TRANsform:TIME:IMPulse:WIDTh?
Or
CALCulate<Ch>:TRACe<Tr>:TRANsform:TIME:IMPulse:WIDTh <time>
CALCulate<Ch>:TRACe<Tr>:TRANsform:TIME:IMPulse:WIDTh?
Sets or reads out the impulse width (time domain transformation resolution), coupled with the Kaiser-Bessel window shape β parameter. The impulse width setting changes the β parameter and setting of β parameter changes the impulse width. |
---|
command/query |
CALCulate<Ch>[:SELected] — active trace of channel <Ch>,
Or
CALCulate<Ch>:TRACe<Tr> — trace <Tr> of channel <Ch>,
<Ch> = {[1]|2|...16}
<Tr> = {[1]|2|...16}
<time> the impulse width, the allowable range varies depending on the specified frequency range
sec (second)
Sets the value of the limit, which is closer to the specified value.
<numeric>
Analysis > Time Domain > Window > Impulse Width
(when the transformation type is set to Bandpass or Lowpass Impulse)
Equivalent COM Command
SCPI.CALCulate(Ch).SELected.TRANsform.TIME.IMPulse.WIDTh
note |
This command is similar to CALCulate<Ch>[:SELected]:TRANsform:TIME:IMPulse:WIDTh only. |
Value = app.SCPI.CALCulate(Ch).SELected.TRANsform.TIME.IMPulse.WIDTh
app.SCPI.CALCulate(Ch).SELected.TRANsform.TIME.IMPulse.WIDTh = 1e–8
Double (read/write)
Back to CALCulate
Back to Command Finder
Back to Command Finder (From User Interface)