Please enable JavaScript to view this site.

 Network Analyzers using S4VNA software

SCPI Command

CALCulate<Ch>[:SELected]:MARKer:DISCrete {OFF|ON|0|1}

CALCulate<Ch>[:SELected]:MARKer:DISCrete?

Or

CALCulate<Ch>:TRACe<Tr>:MARKer:DISCrete {OFF|ON|0|1}

CALCulate<Ch>:TRACe<Tr>:MARKer:DISCrete?

Description

Turns the marker discrete mode ON/OFF.

command/query

Target

All traces of channel <Ch> (if the marker coupling is set to ON by the CALC:MARK:COUP command),

Or

CALCulate<Ch>[:SELected] — active trace of channel <Ch>,

Or

CALCulate<Ch>:TRACe<Tr> — trace <Tr> of channel <Ch>,

<Ch> = {[1]|2|...16}

<Tr> = {[1]|2|...16}

Parameter

{ON|1}

ON

{OFF|0}

OFF

Query Response

{0|1}

Preset Value

1

Equivalent Softkeys

Marker > Properties > Discrete {ON/OFF}


Equivalent COM Command

SCPI.CALCulate(Ch).SELected.MARKer(1).DISCrete


note

This command is similar to CALCulate<Ch>[:SELected]:MARKer:DISCrete only.


Syntax

Status = app.SCPI.CALCulate(Ch).SELected.MARKer.DISCrete

app.SCPI.CALCulate(Ch).SELected.MARKer.DISCrete = false

Type

Boolean (read/write)


Warning

Object MARKer has an index of 1, which can be omitted in Visual Basic, but it cannot be omitted in other programming languages.



 

Back to CALCulate

Back to Command Finder

Back to Command Finder (From User Interface)

 

Rev.:  22.4