 Network Analyzers using S4VNA software

SCPI Command

CALCulate<Ch>:FSIMulator:BALun:DZConversion:BPORt<Bpt>:Z0[:R] <numeric>

CALCulate<Ch>:FSIMulator:BALun:DZConversion:BPORt<Bpt>:Z0[:R]?

Description

Sets or reads out the impedance value for the differential impedance conversion function of the balanced port. The impedance is real. The default impedance value equals to 100 Ω.

command/query

Target

Balanced Port <Bpt> of channel <Ch>,

<Ch>={[1]|2|...16}

<Bpt>={[1]|2}, 1 or 2 for the Bal-Bal topology, always 1 for the SE-Bal, SE-SE-Bal and Bal topology.

Parameter

<numeric> the new value of the differential impedance of the balanced port from 1 mΩ to 10 MΩ.

Unit

Ω (Ohm)

Out of Range

Sets the value of the limit, which is closer to the specified value.

Query Response

<numeric>

Preset Value

100 Ω

Equivalent Softkeys

Analysis > Fixture Simulator > Diff ZConversion > Bal Port n


Equivalent COM Command

SCPI.CALCulate(Ch).FSIMulator.BALun.DZConversion.BPORt(Bpt).Z0.R

Syntax

Status = app.SCPI.CALCulate(Ch).FSIMulator.BALun.DZConversion.BPORt(Bpt).Z0.R

app.SCPI.CALCulate(Ch).FSIMulator.BALun.DZConversion.BPORt(Bpt).Z0.R = 20

Type

Double (read/write)


 

Back to CALCulate

Back to Command Finder

Back to Command Finder (From User Interface)

 