Please enable JavaScript to view this site.

 Network Analyzers using S4VNA software

SCPI Command

CALCulate<Ch>:FSIMulator:BALun:CZConversion:BPORt<Bpt>:Z0[:R] <numeric>

CALCulate<Ch>:FSIMulator:BALun:CZConversion:BPORt<Bpt>:Z0[:R]?

Description

Sets or reads out the impedance value for the common impedance conversion function of the balanced port. The impedance is real. The default impedance value equals to 25 Ω.

command/query

Target

Balanced Port <Bpt> of channel <Ch>,

<Ch>={[1]|2|...16}

<Bpt>={[1]|2}, 1 or 2 for the Bal-Bal topology, always 1 for the SE-Bal, SE-SE-Bal and Bal topology.

Parameter

<numeric> the new value of the common impedance of the balanced port from 1 mΩ to 10 MΩ.

Unit

Ω (Ohm)

Out of Range

Sets the value of the limit, which is closer to the specified value.

Query Response

<numeric>

Preset Value

25 Ω

Equivalent Softkeys

Analysis > Fixture Simulator > Cmn ZConversion > Bal Port n


Equivalent COM Command

SCPI.CALCulate(Ch).FSIMulator.BALun.CZConversion.BPORt(Bpt).Z0.R

Syntax

Value = app.SCPI.CALCulate(Ch).FSIMulator.BALun.CZConversion.BPORt(Bpt).Z0.R

app.SCPI.CALCulate(Ch).FSIMulator.BALun.CZConversion.BPORt(Bpt).Z0.R = 20

Type

Double (read/write)


 

Back to CALCulate

Back to Command Finder

Back to Command Finder (From User Interface)

 

Rev.:  22.4