SCPI Command

SENSe:SWEep:PULSe:MODulator:ENABle {OFF|ON|0|1}

SENSe:SWEep:PULSe:MODulator:ENABle?

Description

Turns the internal modulator in asynchronous mode ON/OFF. In synchronous mode, the modulator is always on. There is a continuous stimulus at the analyzer port if the modulator is turned off.

Note: S5180B model only.

command/query

Target

Analyzer

Parameter

The analyzer port transmits a stimulus:

{ON|1}

modulated by an internal modulator

{OFF|0}

continuous

Query Response

{0|1}

Preset Value

1

Equivalent Softkeys

Stimulus > Pulse Meas > Internal Modulator


Equivalent COM Command

None


 

Back to Pulse Measurement S5180B