Please enable JavaScript to view this site.

 Network Analyzers using S2VNA software

SCPI Command

SENSe<ch>:PLS2:MODE {OFF|PIP|NBW|PROF|HIPR|PTOP}

SENSe<ch>:PLS2:MODE?

Description

Turn pulse measurements on and sets their mode.

Note: S5243 model only.

command/query

Target

Channel <Ch> = {[1]|2| ... 16}

Parameter

<char> defines the mode:

OFF

Pulse mode disabled

PIP

"Point-in-pulse" — frequency response of the DUT is measured using wideband detection

NBW

"Narrowband" — frequency response of the DUT is measured using narrowband detection

PROF

"Pulse profile" — pulse parameters (amplitude envelope and pulse phase) are measured in the time domain at the width of one pulse

HIPR

"Pulse profile high resolution" — pulse profile is measured with higher time resolution by using a series of pulses instead of one pulse

PTOP

"Pulse to pulse" — the drift of DUT parameters is tracked during the action of one burst of pulses with same carrier frequency

Query Response

{OFF|PIP|NBW|PROF|HIPR|PTOP}

Preset Value

OFF

Equivalent Softkeys

Stimulus > Pulse Meas > Pulse Mode


Equivalent COM Command

None


 

Back to Pulse Measurement S5243

Back to Command Finder (From User Interface)

 

Rev.:  24.1