Please enable JavaScript to view this site.

 Network Analyzers using S2VNA software

SCPI Command

SENSe<Ch>:CORRection:COLLect:METHod:TRL:MULTiline[:STATe] {OFF|ON|0|1}

SENSe<Ch>:CORRection:COLLect:METHod:TRL:MULTiline[:STATe]?

Description

Turns the multi-line TRL option ON/OFF. Determines which TRL algorithm is used when calculating the calibration coefficients using the SENS:CORR:COLL:SAVE command. If turned on, the multi-line TRL algorithm is used. If turned off, the classic TRL algorithm is used. Multi-line TRL option is not supported by Planar 304, Planar 304/1 analyzers.

command/query

Target

Channel <Ch>,

<Ch>={[1]|2|...16}

Parameter

{ON|1}

ON

{OFF|0}

OFF

Query Response

{0|1}

Preset Value

0

Related Commands

SENS:CORR:COLL:SAVE

SENS:CORR:COLL:METH:TRL2

Equivalent Softkeys

Calibration > Calibrate > n-Port TRL Cal > Multiline


Equivalent COM Command

None


 

Back to SENSe

Back to Command Finder

 