Please enable JavaScript to view this site.

 Network Analyzers using S2VNA software

SCPI Command

SENSe<Ch>:CORRection:COLLect:SAVE

Description

Calculates the calibration coefficients from the calibration standards measurements depending on the selected calibration type. The calibration type is selected by one of commands SENS:CORR:COLL:METH:XXXX.

On completion of the command, all the calibration standards measurements are cleared, and the error correction automatically turns ON.

At the attempt to execute this command before all the needed standards are measured, an error occurs, and the command is ignored.

no query

Target

Channel <Ch>,

<Ch>={[1]|2|...16}

Related Commands

Calibration type selection:

SENS:CORR:COLL:METH:XXXX

Calibration standards measurement:

SENS:CORR:COLL:ISOL

SENS:CORR:COLL:LOAD

SENS:CORR:COLL:OPEN

SENS:CORR:COLL:SHOR

SENS:CORR:COLL:THRU

SENS:CORR:COLL:TRLT

SENS:CORR:COLL:TRLL

SENS:CORR:COLL:TRLR

Equivalent Softkeys

Calibration > Calibrate > Response (Open) | Respose (Short) | Response (Thru) | 1-port SOL Cal | One Path 2-Port Cal |n-Port SOLT Cal. | n-Port TRL Cal > Apply


Equivalent COM Command

SCPI.SENSe(Ch).CORRection.COLLect.SAVE

Syntax

app.SCPI.SENSe(Ch).CORRection.COLLect.SAVE

Type

Method


 

Back to SENSe

Back to Command Finder

 

Rev.:  24.2