Please enable JavaScript to view this site.

 Network Analyzers using S2VNA software

SCPI Command

SENSe<Ch>:CORRection:COLLect:METHod:SOLT2 <port1>,<port2>

Description

Selects the port and sets the full two-port (SOLT) calibration type for the calculation of the calibration coefficients on completion of the calibration executed by the SENS:CORR:COLL:SAVE command.

no query

Target

Channel <Ch>,

<Ch>={[1]|2|...16}

Parameter

<port1>

The first port number from 1 to 2

<port2>

The second port number from 1 to 2

Out of Range

If the same port numbers are specified, an error occurs.

Related Commands

SENS:CORR:COLL:SAVE

Equivalent Softkeys

Calibration > Calibrate > 2–Port SOLT Cal


Equivalent COM Command

SCPI.SENSe(Ch).CORRection.COLLect.METHod.SOLT2

Syntax

app.SCPI.SENSe(Ch).CORRection.COLLect.METHod.SOLT2 = Array(2, 1)

Type

Variant (array of Long) (write only)


 

Back to SENSe

Back to Command Finder

 

Rev.:  24.2