SCPI Command
SENSe:CORRection:COLLect:CKIT:STAN<Std>:C3 <numeric>
SENSe:CORRection:COLLect:CKIT:STAN<Std>:C3?
Sets or reads out the C3 value for the open calibration standard. |
---|
command/query |
Standard <Std> of the calibration kit,
<Std>={[1]|2|...N}, where N — the number of the standards in the calibration kit
<numeric> the arbitrary impedance value from –1E18 to 1E18
1E–45 F/Hz3 (Farad/Hertz3)
Out of Range
Sets the value of the limit, which is closer to the specified value.
<numeric>
Calibration > Cal Kit > Define STDs -> Open n -> C3 10–45 F/Hz3
Equivalent COM Command
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).C3
Value = app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).C3
app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).C3 = 100
Double (read/write)