SCPI Command
SENSe:CORRection:COLLect:CKIT:STAN<Std>:C2 <numeric>
SENSe:CORRection:COLLect:CKIT:STAN<Std>:C2?
Sets or reads out the C2 value for the open calibration standard. |
---|
command/query |
Standard <Std> of the calibration kit,
<Std>={[1]|2|...N}, where N — the number of the standards in the calibration kit
<numeric> the arbitrary impedance value from –1E18 to 1E18
1E–36 F/Hz2 (Farad/Hertz2)
Out of Range
Sets the value of the limit, which is closer to the specified value.
<numeric>
Calibration > Cal Kit > Define STDs -> Open n -> C2 10–36 F/Hz2
Equivalent COM Command
SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).C2
Value = app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).C2
app.SCPI.SENSe(Ch).CORRection.COLLect.CKIT.STAN(Std).C2 = 100
Double (read/write)
Back to SENSe
Back to Command Finder