A marker is a tool for numerical readout of a stimulus value and value of the measured parameter in a specific point on the trace. Up to 16 markers can be activated on each trace. A trace with two markers is shown in the figure below.
The markers allow to perform the following tasks:
•Reading absolute values of a stimulus and a measured parameter in selected points on the trace.
•Reading relative values of a stimulus and a measured parameter related to the reference point.
•Search for specific points on the trace (minimum, maximum, target level, etc.).
•Determining trace parameters (statistics, bandwidth, etc.).
•Editing stimulus parameters using markers.
Markers can have the following indicators:
1 ∇ |
Symbol and number of the active marker on a trace. |
Δ 2 |
Symbol and number of the inactive marker on a trace. |
Symbol of the active marker on a stimulus axis. |
|
Δ |
Symbol of the inactive marker on a stimulus axis. |
The marker data field contains the marker number, stimulus value, and the measured parameter value. The number of the active marker is highlighted in an inverse color.
The marker data field contents vary depending on the display format (rectangular or circular):
•In rectangular format, the marker shows the measurement parameter value plotted along Y-axis in the active format (See the table below).
Format Type Description |
Label |
Data Type (Y-axis) |
Measurement Unit (Y-axis) |
---|---|---|---|
Logarithmic Magnitude |
Log Mag |
S-parameter logarithmic magnitude: , |
Decibel (dB) |
Voltage Standing Wave Ratio |
SWR |
|
Dimensionless value |
Phase |
Phase |
S-parameter phase from –180° to +180°:
|
Degree (°) |
Expanded Phase |
Expand Phase |
S-parameter phase, measurement range expanded to from below –180° to over +180° |
Degree (°) |
Group Delay |
Group Delay |
Signal propagation delay within the DUT: ,, |
Second (sec.) |
Linear Magnitude |
Lin Mag |
S-parameter linear magnitude: |
Dimensionless value |
Real Part |
Real |
S-parameter real part: |
Dimensionless value |
Imaginary Part |
Imag |
S-parameter imaginary part: |
Dimensionless value |
•In circular format, the marker shows two or three values listed in the table below.
Label |
Marker Readings (Measurement Unit) |
||
---|---|---|---|
Reading 1 |
Reading 2 |
Reading 3 |
|
Smith (Lin) |
Linear magnitude |
Phase (°) |
— |
Smith (Log) |
Logarithmic magnitude (dB) |
Phase (°) |
— |
Smith (Re/Im) |
Real part |
Imaginary part |
— |
Smith (R + jX) |
Resistance (Ω) |
Reactance (Ω) |
Equivalent capacitance or inductance (F/H) |
Smith (G + jB) |
Conductance (S) |
Susceptance (S) |
Equivalent capacitance or inductance (F/H) |
Polar (Lin) |
Linear magnitude |
Phase (°) |
— |
Polar (Log) |
Logarithmic magnitude (dB) |
Phase (°) |
— |
Polar (Re/Im) |
Real part |
Imaginary part |
— |
Marker Addition
To enable a new marker, use the following softkeys: Markers > Add Marker |
|
|
note |
The new marker appears as the active marker in the middle of the stimulus axis. The input field for the marker stimulus value is activated. |
Marker Deletion
To delete a marker, use the following softkeys: Markers > Remove Marker To delete all the markers, use the following softkeys: Markers > Remove All Markers |
To activate a marker by its number, use the following softkeys: Markers > Select > Marker n |
|
To activate a marker from the list of markers, use the following softkeys: Markers > Select Next |
|
|
note |
A marker can be activated by clicking on it. |
The active marker must be selected before setting the marker stimulus value. The stimulus value must be set by entering the numerical value from the keyboard, by arrows, by dragging the marker using the mouse (See Marker Stimulus Value Setting), or by enabling the search function (See Marker Position Search Functions).
To set the marker stimulus value, use the following softkeys: Markers > Edit Stimulus or click on the stimulus value field using the mouse. Then, enter the value using the numerical keys on the keypad, by «↑», «↓» arrows. |
|
|