 Network Analyzers using RVNA and RNVNA software

General use and features:

•System Z0 should be set to 1 Ω before calibration. Offset Z0 and terminal impedance in the calibration standard definition also should be set to 1 Ω (See System Impedance Z0).

•Waveguide calibration uses two offset short standards instead of a combination of short and open standards. Typically, 1/8λ0 and 3/8λ0 offset short standards are used, where λ0 — wave length in waveguide at the mean frequency.


In waveguide calibration, one of the offset short standards is assigned as the open standard. Therefore, the GUI contains 3/8 Offset Short and 1/8 Offset Short softkey.

To perform calibration, connect the appropriate waveguide to the analyzer port and press the desired softkey.

Waveguide Calibration