General use and features:

System Z0 should be set to 1 Ω before calibration. Offset Z0 and terminal impedance in the calibration standard definition also should be set to 1 Ω (See System Impedance Z0).

Waveguide calibration uses two offset short standards instead of a combination of short and open standards. Typically, 1/8λ0 and 3/8λ0 offset short standards are used, where λ0 wave length in waveguide at the mean frequency.


In waveguide calibration, one of the offset short standards is assigned as the open standard. Therefore, the GUI contains 3/8 Offset Short and 1/8 Offset Short softkey.

To perform calibration, connect the appropriate waveguide to the analyzer port and press the desired softkey.

Waveguide Calibration